Bridge Analyzers home

Bridge Analyzers, Inc.
1805-B Clement Ave.
Alameda, Calif. 94501

Phone: (510) 337-1605
Fax: (510) 337-0388



Model 900141 Lo-Ox/Hi-Ox CO2/O2 MAP Headspace Gas Analyzer

Enhanced Oxygen measurement from 100% down to 0.001%

  • Simultaneously Measures CO2 and O2 Headspace Gas.
  • Industrial NEMA-4 Design - Sealed for Harsh Environments.
  • Internal Battery for Standalone Use.
  • Simple and fast – Single button - Headspace gases in 15 Seconds
  • Infrared for CO2, Electro-Chemical with 2-Point Span for O2.
  • No Need for Frequent Recalibration– Stable, Long-Lasting Accuracy
  • Small (6”x3”x3”), lightweight (2.5 Lbs), and rugged.

Model 900141 MAP Headspace Gas Analyzer Specifications

Gasses Measured: Carbon Dioxide (CO2) and Oxygen (O2)
Measurement Methods: CO2: NDIR (Non-Dispersive Infra-Red.)
O2: Electro Chemical Sensor
Measuring Ranges: CO2: 00.00-100.0%
O2: 00.000-100.0%
Resolution: CO2: 0.01% Below 10.0%, 0.1% above 10.0%,
O2: 0.001% Below 2.00%, 0.01% above 2.00%
Accuracy:
Repeatability:
4% relative to gas reading.
1% relative to gas reading.
Response time: 8 Seconds to 95% Final Reading,
15 Seconds to 99.5% Final Reading.
Sampling Rate: 4 ml/sec
Sample Time Adjustable in 1 second increments. Factory default 15 Seconds.
Gas-borne Contaminate Protection: External and internal water blocking and particle filters protect analyzer from test gas-borne contamination.
Warm-up time: Full accuracy in 3 minutes.
Operating Temperature: 0 to +40 degrees C.
Power:
Line Power
Internal Battery

110VAC / 220VAC Universal Wall Supply
NiMH battery pack, 6-8 hours operation per charge.
Battery Charger/ External Supply: 110 VAC/60 Hz /220 VAC/50 Hz Regulated ‘E-Star’ compliant power supply/charger.
Weight:
Unit Alone
In Soft Case

2.5 Lbs.
3.5 Lbs.
Size:
Unit Alone
In Soft Case

6 ¼ L x 3W x 3 ½H
8 ½L x 5W x 5H
PC Interface: Fully Sealed Serial Port, 9600 baud.

Bridge Analyzers, Inc.  1805-B Clement Ave.  Alameda, Calif. 94501  510.337.1605